USP Equivalency Study: PathoSEEK® Salmonella & E. coli for Dietary Supplements
An independent contract research laboratory compared the PathoSEEK® Salmonella and E. coli Multiplex Detection Assay with MaGiC Lysis against USP <2022> compendial plating in six dietary supplement matrices. The assay met all six USP <1223> parameters for qualitative methods: specificity, limit of detection, repeatability, ruggedness, robustness, and equivalency.
The qPCR method agreed with compendial results on every inoculated sample, at every inoculation level, on both qPCR instruments, with zero false positives. Non-inferiority was demonstrated at low, intermediate, and high levels (no discordant pairs; δ = 0.20, α = 0.05). A separate dilution series using certified reference materials put the LOD₅₀ at 0.23 to 0.38 CFU per test portion.
There were no discordant results to explain. The one practical note concerns botanical powder that has been autoclaved, covered in the Robustness section.
Study at a glance
| Report | USP <1223> Alternative Method Equivalency Study for Detection of Salmonella spp. and Escherichia coli in Dietary Supplements (P0103-RPT-002), Version 1 (October 2026) |
|---|---|
| Method | PathoSEEK® Salmonella and Escherichia coli Multiplex Detection Assay with MaGiC Lysis |
| Target organisms | Salmonella enterica NCTC 6676, Escherichia coli NCTC 12923 / ATCC 8739 |
| Co-inoculated organism | Staphylococcus aureus NCTC 10788 / ATCC 6538 (part of a companion triplex study; not a target of this assay) |
| Off-target challenge | Klebsiella oxytoca NCTC 8167, Citrobacter koseri ATCC 25408 |
| Matrices | Energy shot, vitamin gummy, vitamin powder, botanical powder, MCT oil, chocolate bar |
| Reference method | USP <2022> (harmonized with USP <62>): MacConkey Agar for E. coli, Xylose Lysine Deoxycholate (XLD) Agar for Salmonella |
| Confirmation | Indole and Citrate tests on recovered colonies |
| LOD reference material | Certified reference materials (CRMs), matrix-independent dilution series |
| Instruments | Bio-Rad CFX Opus, BioMolecular Systems Mic, Myra liquid handler for automated extraction |
| Conducted by | Cambium Analytica |
Results by matrix
All six matrices showed complete agreement with compendial methods at every level tested. Full data for each parameter follows below.
| Matrix | Low‑level spike, CFU/test portion (Salmonella / E. coli) | Low‑level detection, qPCR vs. compendial | Intermediate and high‑level detection | Specificity | Highest average RSD (Cq) | Relative trueness |
|---|---|---|---|---|---|---|
| Energy shot | 2 / 8 | 8/8 vs. 8/8 | 100% | 100% | 0.61% | 100% |
| Vitamin gummy | 2 / 5 | 8/8 vs. 8/8 | 100% | 100% | 5.00% | 100% |
| Vitamin powder | 4 / 2 | 8/8 vs. 8/8 | 100% | 100% | 1.56% | 100% |
| Botanical powder | 1 / 2 | 8/8 vs. 8/8 | 100% | 100% | 1.87% | 100% |
| MCT oil | 6 / 1 | 8/8 vs. 8/8 | 100% | 100% | 2.05% | 100% |
| Chocolate bar | 1 / 3 | 8/8 vs. 8/8 | 100% | 100% | 1.67% | 100% |
Low-level detection counts apply to each organism on each instrument. Source: Tables 3, 6 to 9, 12, and 15 of P0103-RPT-002.
Chocolate is the most demanding matrix in the set, with high fat, cocoa polyphenols, and methylxanthines. Per the User Guide, its lysate is diluted 1:100 before qPCR and its internal control is added at 1:5,000. With those steps, the internal amplification control amplified in every chocolate sample, and every low-level sample was detected, including those carrying 1 CFU of Salmonella per test portion.
Study design
Each sample was tested side by side using the qPCR method and USP <2022> compendial plating, with compendial colonies confirmed by biochemical tests. Before the matrix work, a matrix-independent LOD was determined using certified reference materials.
| Inoculation level | Samples per matrix | Purpose |
|---|---|---|
| Uninoculated blank (0 CFU) | 2 | Confirm no false positives from matrix |
| Off‑target (K. oxytoca + C. koseri) | 2 | Confirm no cross-reactivity |
| Low level, near limit of detection | 8 | Challenge sensitivity |
| Intermediate level | 8 | Confirm detection at moderate contamination |
| High level | 2 | Confirm complete detection |
Every positive sample was co-inoculated with Salmonella, E. coli, and S. aureus, so each channel had to find its target in a mixed contamination. Samples were enriched in buffered peptone water for 20 to 24 hours at 37 °C. DNA was extracted with MaGiC Lysis on the Myra automated liquid handler, and each extract was run on both the Bio-Rad CFX Opus and the Mic. Each reaction detected E. coli (FAM), Salmonella (ROX), and an internal amplification control (HEX) in a single well.
The CRM series ran at 5, 2.5, 1, 0.5, 0.25, and 0.1 CFU per test portion, five replicates per level, extracted directly without enrichment.
Specificity
The assay produced zero false positives across all six matrices and both instruments. Specificity was 100% for the qPCR method and the compendial method alike.
Acceptance criterion: no false-positive detections in uninoculated blanks or samples spiked with off-target organisms (K. oxytoca NCTC 8167, C. koseri ATCC 25408); specificity = 100%.
| Matrix | Instrument | True negatives | False positives | Specificity, qPCR | Specificity, compendial |
|---|---|---|---|---|---|
| Energy shot | Bio‑Rad / Mic | 8 / 8 | 0 / 0 | 100% | 100% |
| Vitamin gummy | Bio‑Rad / Mic | 8 / 8 | 0 / 0 | 100% | 100% |
| Vitamin powder | Bio‑Rad / Mic | 8 / 8 | 0 / 0 | 100% | 100% |
| Botanical powder | Bio‑Rad / Mic | 8 / 8 | 0 / 0 | 100% | 100% |
| MCT oil | Bio‑Rad / Mic | 8 / 8 | 0 / 0 | 100% | 100% |
| Chocolate bar | Bio‑Rad / Mic | 8 / 8 | 0 / 0 | 100% | 100% |
True negatives are counted per target channel (4 samples × 2 targets). Source: Table 3. The internal amplification control amplified at Cq ≤ 35 in every sample, confirming the negatives were valid. Inclusivity and exclusivity across strain panels were established separately by Medicinal Genomics and were not re‑tested in this study.
Limit of detection
The CRM dilution series put the LOD₅₀ at 0.23 to 0.38 CFU per test portion, and in-matrix detection at the low level showed no statistically significant difference from compendial plating in any matrix. LOD was evaluated two ways: a matrix-independent CRM dilution series, and Probability of Detection (POD) analysis in each matrix.
Acceptance criteria: LOD₅₀ at or below about 1 CFU per test portion for each target. In matrix, the 95% confidence interval for dPOD (qPCR POD minus compendial POD) must include zero at each level tested.
| Nominal CFU/test portion | E. coli, Bio‑Rad | E. coli, Mic | Salmonella, Bio‑Rad | Salmonella, Mic |
|---|---|---|---|---|
| 5 CFU/test portion | 5/5 | 5/5 | 5/5 | 5/5 |
| 2.5 CFU/test portion | 5/5 | 5/5 | 5/5 | 5/5 |
| 1 CFU/test portion | 5/5 | 5/5 | 5/5 | 5/5 |
| 0.5 CFU/test portion | 4/5 | 4/5 | 3/5 | 3/5 |
| 0.25 CFU/test portion | 2/5 | 2/5 | 2/5 | 1/5 |
| 0.1 CFU/test portion | 0/5 | 0/5 | 0/5 | 0/5 |
| Target | Instrument | LOD50 (95% CI) | LOD95 (95% CI) |
|---|---|---|---|
| E. coli | Bio‑Rad CFX Opus | 0.23 (0.12, 0.43) | 0.99 (0.53, 1.87) |
| E. coli | Mic | 0.27 (0.14, 0.51) | 1.16 (0.62, 2.19) |
| Salmonella spp. | Bio‑Rad CFX Opus | 0.33 (0.17, 0.61) | 1.41 (0.75, 2.65) |
| Salmonella spp. | Mic | 0.38 (0.20, 0.71) | 1.63 (0.87, 3.05) |
Nominal CFU per test portion, positive = target Cq ≤ 37. Estimates computed with the PODLOD program referenced in ISO 16140‑2. Source: Tables 4 and 5.
| Matrix | Organism | CFU/test portion | POD, qPCR | POD, compendial | dPOD (95% CI) |
|---|---|---|---|---|---|
| Energy shot / Salmonella spp. | Salmonella spp. | 2 | 1.00 | 1.00 | 0.00 (−0.31, 0.31) |
| Energy shot / E. coli | E. coli | 8 | 1.00 | 1.00 | 0.00 (−0.31, 0.31) |
| Vitamin gummy / Salmonella spp. | Salmonella spp. | 2 | 1.00 | 1.00 | 0.00 (−0.31, 0.31) |
| Vitamin gummy / E. coli | E. coli | 5 | 1.00 | 1.00 | 0.00 (−0.31, 0.31) |
| Vitamin powder / Salmonella spp. | Salmonella spp. | 4 | 1.00 | 1.00 | 0.00 (−0.31, 0.31) |
| Vitamin powder / E. coli | E. coli | 2 | 1.00 | 1.00 | 0.00 (−0.31, 0.31) |
| Botanical powder / Salmonella spp. | Salmonella spp. | 1 | 1.00 | 1.00 | 0.00 (−0.31, 0.31) |
| Botanical powder / E. coli | E. coli | 2 | 1.00 | 1.00 | 0.00 (−0.31, 0.31) |
| MCT oil / Salmonella spp. | Salmonella spp. | 6 | 1.00 | 1.00 | 0.00 (−0.31, 0.31) |
| MCT oil / E. coli | E. coli | 1 | 1.00 | 1.00 | 0.00 (−0.31, 0.31) |
| Chocolate bar / Salmonella spp. | Salmonella spp. | 1 | 1.00 | 1.00 | 0.00 (−0.31, 0.31) |
| Chocolate bar / E. coli | E. coli | 3 | 1.00 | 1.00 | 0.00 (−0.31, 0.31) |
Low inoculation level, n = 8 per row. Results were identical on the Bio‑Rad CFX Opus and the Mic. At 0 CFU every POD was 0.00; at the intermediate and high levels every POD was 1.00. Source: Tables 6 to 9.
At the low level, compendial plating recovered 1 to 8 CFU per test portion, and the qPCR method detected every sample for both targets in every matrix. That matches the CRM results, so none of the six matrices measurably reduced sensitivity. Using the MPN table in ISO 16140-3, the report estimates an in-matrix LOD₅₀ below 1.0 CFU per test portion for every matrix on both instruments (Tables 10 and 11).
Repeatability
All average RSDs of log-transformed Cq values were at or below 5.00%, and no individual sample exceeded 6.09%. Acceptance criterion: RSD ≤10%.
RSD was measured on the internal amplification control (HEX), which is present in every valid sample. Each DNA extract was run on both the Bio-Rad CFX Opus and the Mic, and RSD was calculated for each sample from its two readings.
| Matrix | Highest average RSD | Highest individual RSD, inoculated samples | Overall average RSD (n = 22) |
|---|---|---|---|
| Energy shot | 0.61% | 2.83% | 0.46% |
| Vitamin gummy | 5.00% | 6.09% | 4.03% |
| Vitamin powder | 1.56% | 1.98% | 1.41% |
| Botanical powder | 1.87% | 2.54% | 1.68% |
| MCT oil | 2.05% | 2.79% | 1.75% |
| Chocolate bar | 1.67% | 1.80% | 1.40% |
Source: Table 12
The report gives averages only for uninoculated samples. The Mic generally read HEX about 2 cycles earlier than the Bio-Rad. The largest gap was in vitamin gummy (4.87 cycles on average), which accounts for its higher RSD; the offset did not change with inoculation level and did not affect any detect/non-detect result.
Ruggedness
The qPCR method returned identical detect/non-detect results on both instruments for every sample and both targets, in every matrix: 264 paired determinations (132 samples × 2 targets), 100% concordant. Ruggedness under USP <1223> measures whether qualitative results hold when operating conditions change.
In this study the varied condition was the qPCR instrument. All extractions were run on the Myra automated liquid handler. Every sample detected on one instrument was detected on the other, every uninoculated sample was negative on both, and results on both instruments matched the compendial method.
Robustness
The internal amplification control amplified in every sample across all six matrices, both instruments, and all inoculation levels. No systematic PCR inhibition was observed.
| Matrix | Challenge it presents |
|---|---|
| Energy shot | Aqueous liquid with organic acids, vitamins, caffeine |
| Vitamin gummy | Semi‑solid gelatin with sugars, colorants, vitamins |
| Vitamin powder | Fine powder with mineral and vitamin components |
| Botanical powder | Plant matrix with polyphenols, tannins, and other potential PCR inhibitors |
| MCT oil | Hydrophobic lipid |
| Chocolate bar | High‑fat confection with cocoa polyphenols, methylxanthines, and sugars |
Sample prep followed the PathoSEEK® User Guide, including pre-qPCR lysate dilutions for botanical powder (1:10), energy shot (1:10), and chocolate (1:100), and a 1:5,000 internal control dilution for chocolate.
Note on heat-treated botanicals
The first botanical powder lot tested had been autoclaved to lower its native bacterial load. It gave invalid results in the pathogen channels, most likely because autoclaving leaves a large amount of cell-free DNA that competes in the reaction. Re-extracting those samples with the optional Grim Reefer free DNA removal step from the User Guide gave valid results with no inhibition.
The botanical powder data on this page come from a new, non-autoclaved lot. For heat-treated, autoclaved, or irradiated material, the report recommends adding Grim Reefer or an equivalent DNA cleanup before lysis, and verifying low-level detection before routine use.
Equivalency
Non-inferiority to compendial plating was demonstrated at all three inoculation levels on both instruments using USP <1223> Approach 1 (paired samples, one-sided Z-test, δ = 0.20, α = 0.05).
| Instrument | Level | Paired observations | Positive rate, qPCR | Positive rate, compendial | Z‑statistic | Critical value | Result |
|---|---|---|---|---|---|---|---|
| Bio‑Rad CFX Opus / Low | Low | 96 | 1.000 | 1.000 | ∞ (no discordant pairs) | 1.645 | ✓ PASS |
| Bio‑Rad CFX Opus / Intermediate | Intermediate | 96 | 1.000 | 1.000 | ∞ (no discordant pairs) | 1.645 | ✓ PASS |
| Bio‑Rad CFX Opus / High | High | 24 | 1.000 | 1.000 | ∞ (no discordant pairs) | 1.645 | ✓ PASS |
| Mic / Low | Low | 96 | 1.000 | 1.000 | ∞ (no discordant pairs) | 1.645 | ✓ PASS |
| Mic / Intermediate | Intermediate | 96 | 1.000 | 1.000 | ∞ (no discordant pairs) | 1.645 | ✓ PASS |
| Mic / High | High | 24 | 1.000 | 1.000 | ∞ (no discordant pairs) | 1.645 | ✓ PASS |
Instruments were analyzed separately because they ran the same extracts. Source: Tables 13 and 14.
| Metric | Low level | Intermediate level | High level | Overall |
|---|---|---|---|---|
| Sensitivity, qPCR | 100% | 100% | 100% | 100% |
| Sensitivity, compendial | 100% | 100% | 100% | 100% |
| Relative trueness | 100% | 100% | 100% | 100% |
| False positive rate | 0% | 0% | 0% | 0% |
Source: Tables 15 and 16.
Conclusion
The PathoSEEK® Salmonella and E. coli Multiplex Detection Assay with MaGiC Lysis met all six USP <1223> parameters and demonstrated equivalency to USP compendial plating for qualitative detection of Salmonella spp. and E. coli in energy shot, vitamin gummy, vitamin powder, botanical powder, MCT oil, and chocolate matrices.
Related Resources:
The PathoSEEK® Salmonella and E. coli Multiplex Detection Assay with MaGiC Lysis met all six USP <1223> parameters and demonstrated equivalency to USP compendial plating for qualitative detection of Salmonella spp. and E. coli in energy shot, vitamin gummy, vitamin powder, botanical powder, MCT oil, and chocolate matrices.
Frequently asked questions
Is PathoSEEK® equivalent to USP / plating for Salmonella and E. coli in dietary supplements?
In this study, the assay demonstrated non-inferiority to USP <2022> compendial plating (harmonized with USP <62>) at low, intermediate, and high inoculation levels across six supplement matrices, evaluated against USP <1223>. The two methods agreed on every sample.
Which supplement matrices were tested?
Energy shot, vitamin gummy, vitamin powder, botanical powder, MCT oil, and chocolate bar. They span aqueous liquids, gelatin semi-solids, powders, plant material, oils, and a high-fat confection.
What is the limit of detection?
Using certified reference materials without a matrix, LOD₅₀ ranged from 0.23 to 0.38 CFU per test portion and LOD₉₅ from 0.99 to 1.63, depending on target and instrument. In the matrices, every low-level sample was detected, with compendial plating recovering 1 to 8 CFU per test portion.
Does the assay pick up both organisms when they are present together?
Yes. Every positive sample carried Salmonella, E. coli, and S. aureus at the same time, and each channel detected its target at the low level with no loss of sensitivity.
Does it work on different qPCR instruments?
Yes. Each extract was run on both the Bio-Rad CFX Opus and the BioMolecular Systems Mic, and detect/non-detect results matched on every sample. All extractions used the Myra automated liquid handler.
Can I use it on autoclaved or irradiated botanicals?
The study found that autoclaved botanical powder can carry enough cell-free DNA to interfere with the reaction, and adding the Grim Reefer free DNA removal step resolved it. The report recommends that step, or an equivalent cleanup, for heat-treated or irradiated material, along with in-house verification at low inoculation levels.
If you have questions, please complete our “Contact Us” form and a Medicinal Genomics representative will contact you with details about how to get started with PathoSEEK® for Dietary Supplements.