USP Equivalency Study: PathoSEEK®
Staph & Pseudomonas for Dietary Supplements
An independent contract research laboratory compared the PathoSEEK® Staph & Pseudomonas Multiplex Detection Assay v2 with MaGiC Lysis against USP <2022> compendial plating in five dietary supplement matrices. The study reports results for all six USP <1223> parameters for qualitative methods: specificity, limit of detection, repeatability, ruggedness, robustness, and equivalency. Equivalency, the one parameter tested against a pass/fail criterion, was met.
Non-inferiority to compendial plating was demonstrated at both inoculation levels tested, using 50 independent test portions per level (Z = 5.77 low, perfect agreement high; δ = 0.20, α = 0.05). The qPCR method produced zero false positives, and 98 of 100 inoculated test portions agreed with compendial results.
The two differences came from Botanical Powder at the low inoculation level, where 2 of 10 test portions carrying 1 to 4 CFU per test portion were not detected for either organism. At 10 samples per matrix, the report treats that result as inconclusive. Details are in the Limit of detection section.
Study at a glance
| Report | USP <1223> Alternative Method Equivalency Study for Detection of Staphylococcus aureus and Pseudomonas aeruginosa in Dietary Supplements (P0045-RPT-001-04), Version 4 (October 2026) |
| Method | PathoSEEK® P. aeruginosa & S. aureus Multiplex Detection Assay v2 with MaGiC Lysis |
| Target organisms | Staphylococcus aureus ATCC 6538, Pseudomonas aeruginosa ATCC 9027 |
| Off-target challenge | Salmonella enterica ATCC 6676, Escherichia coli ATCC 11775 |
| Matrices | Energy shot, vitamin gummy, vitamin powder, botanical powder, MCT oil |
| Reference method | USP <2022> (harmonized with USP <62>): Mannitol Salt Agar for S. aureus, Cetrimide Agar for P. aeruginosa |
| Confirmation | DNase Agar and Coagulase Test (S. aureus); Pseudomonas P Agar and Oxidase Test (P. aeruginosa) |
| Conditions varied | Two analysts, two extraction methods (manual and automated), two qPCR instruments |
| Instruments | Bio-Rad CFX Opus, BioMolecular Systems Mic, Myra liquid handler for automated extraction |
| Conducted by | Cambium Analytica |
Results by matrix
Four of five matrices showed complete agreement with compendial methods at every level tested. Full data for each parameter follows below.
| Matrix | Low-level spike, CFU/test portion (P. aeruginosa / S. aureus) | Low-level detection, qPCR vs. compendial | High-level detection | Specificity | Highest average RSD (Cq) | Relative trueness |
|---|---|---|---|---|---|---|
| Energy shot | 21 / 21 | 10/10 vs. 10/10 | 100% | 100% | 3.74% | 100% |
| Vitamin gummy | 14 / 3 | 10/10 vs. 10/10 | 100% | 100% | 2.97% | 100% |
| Vitamin powder | 5 / 4 | 10/10 vs. 10/10 | 100% | 100% | 2.69% | 100% |
| MCT oil | 3 / 2 | 10/10 vs. 10/10 | 100% | 100% | 4.15% | 100% |
| Botanical powder | 4 / 2 | 8/10 vs. 10/10 | 100% | 100% | 4.48% | 93.3% |
Low-level detection counts apply to each organism on each instrument. Relative trueness is measurement-level (six qPCR results per test portion). Source: Tables 2 to 7 and 11 of P0045-RPT-001-04.
Botanical powder is the most demanding matrix in the set, with polyphenols and tannins that can inhibit PCR. The internal amplification control amplified in every botanical sample. The two low-level non-detections carried 1 CFU (S. aureus) and 4 CFU (P. aeruginosa) per test portion by compendial plating, and both organisms were missed in the same two test portions.
Study design
Each sample was tested side by side using the qPCR method and USP <2022> compendial plating, with compendial colonies confirmed by biochemical tests.
Inoculation level | Samples per matrix | Purpose |
|---|---|---|
Uninoculated blank (0 CFU) | 5 | Confirm no false positives from matrix |
Off-target (S. enterica + E. coli) | 5 | Confirm no cross-reactivity |
Low level, 1 to 10 CFU per test portion | 10 | Challenge sensitivity |
High level | 10 | Confirm complete detection |
Both target organisms were co-inoculated into every positive test portion. The low level was set within 1 to 10 CFU per test portion and was not adjusted to produce fractional results.
qPCR testing varied three factors: two analysts, two extraction methods (manual and automated on Myra), and two thermocyclers (Bio-Rad CFX Opus and Mic). Each test portion ran under three extraction and analyst conditions on both instruments, giving six qPCR results per portion against one compendial result. DNA was extracted from enriched samples with MaGiC Lysis. Each reaction detected S. aureus (FAM), P. aeruginosa (Cy5), and an internal amplification control (HEX) in a single well.
Specificity
The assay produced zero false positives across all five matrices, both extraction methods, and both instruments. Specificity was 100% for the qPCR method and the compendial method alike.
Specificity here means no positive results in uninoculated blanks or in samples spiked with off-target organisms (S. enterica ATCC 6676, E. coli ATCC 11775). Recovery of the target organisms is covered under Limit of detection.
Matrix | Extraction | True negatives | False positives | Specificity, qPCR | Specificity, compendial |
|---|---|---|---|---|---|
Energy shot | Manual / Automated | 10 / 10 | 0 / 0 | 100% | 100% |
Vitamin gummy | Manual / Automated | 10 / 10 | 0 / 0 | 100% | 100% |
Vitamin powder | Manual / Automated | 10 / 10 | 0 / 0 | 100% | 100% |
Botanical powder | Manual / Automated | 10 / 10 | 0 / 0 | 100% | 100% |
MCT oil | Manual / Automated | 10 / 10 | 0 / 0 | 100% | 100% |
Source: Table 2
Limit of detection
Low-level detection matched compendial plating in four of five matrices. In Botanical Powder, the qPCR method detected 8 of 10 low-level test portions versus 10 of 10 by plating, and the report treats that difference as inconclusive at this sample size. Per-matrix Probability of Detection (POD) and dPOD (qPCR POD minus compendial POD) are reported descriptively; at 10 samples per level they cannot establish non-inferiority on their own, so the method comparison rests on the pooled analysis under Equivalency.
Matrix | Organism | CFU/test portion | POD, qPCR | POD, compendial | dPOD (95% CI) |
|---|---|---|---|---|---|
Energy shot | P. aeruginosa | 21 | 1.00 | 1.00 | 0.00 (−0.25, 0.25) |
Energy shot | S. aureus | 21 | 1.00 | 1.00 | 0.00 (−0.25, 0.25) |
Vitamin gummy | P. aeruginosa | 14 | 1.00 | 1.00 | 0.00 (−0.25, 0.25) |
Vitamin gummy | S. aureus | 3 | 1.00 | 1.00 | 0.00 (−0.25, 0.25) |
Vitamin powder | P. aeruginosa | 5 | 1.00 | 1.00 | 0.00 (−0.25, 0.25) |
Vitamin powder | S. aureus | 4 | 1.00 | 1.00 | 0.00 (−0.25, 0.25) |
MCT oil | P. aeruginosa | 3 | 1.00 | 1.00 | 0.00 (−0.25, 0.25) |
MCT oil | S. aureus | 2 | 1.00 | 1.00 | 0.00 (−0.25, 0.25) |
Botanical powder | P. aeruginosa | 4 | 0.80 | 1.00 | −0.20 (−0.54, 0.14) |
Botanical powder | S. aureus | 2 | 0.80 | 1.00 | −0.20 (−0.54, 0.14) |
Low inoculation level, n = 10 per row. Results were identical on the Bio-Rad CFX Opus and the Mic. At 0 CFU every POD was 0.00; at the high level every POD was 1.00. Source: Tables 3 to 6.
In Botanical Powder, both organisms were missed in the same two test portions (1 CFU S. aureus and 4 CFU P. aeruginosa recovered by plating), under every analyst, extraction, and instrument condition. The internal amplification control amplified normally in both, so the reaction itself ran. The report notes that a sample-level matrix effect cannot be ruled out.
This study confirmed detection performance at the levels tested. A formal per-matrix LOD₅₀ or LOD₉₅ determination was outside its scope.
Repeatability
Qualitative results agreed across replicate samples in every matrix and level except Botanical Powder at the low level, where 8 of 10 were detected. As supporting data, every average RSD of log-transformed internal control Cq values was below 5%.
RSD was measured on the internal amplification control (HEX), which is present in every valid sample. Six paired comparisons per matrix covered analyst (A1 vs. A2), extraction (manual vs. automated), and instrument (Bio-Rad vs. Mic).
Matrix | Highest average RSD | Highest individual RSD |
|---|---|---|
Energy shot | 3.74% | 6.45% |
Vitamin gummy | 2.97% | 7.72% |
Vitamin powder | 2.69% | 6.29% |
Botanical powder | 4.48% | 10.26% |
MCT oil | 4.15% | 5.80% |
Source: Table 7. One individual sample (Botanical Powder, low level, manual extraction, instrument comparison) reached 10.26%; every other value was below 10%. RSD patterns matched across the Bio-Rad and Mic for each condition.
Ruggedness
The qPCR method returned identical detect/non-detect results across both analysts, both extraction methods, and both instruments, in every matrix. Ruggedness under USP <1223> looks at whether results from the same samples hold up under typical changes in test conditions.
At the high level, concordance was 100% across all factors and matrices. At the low level, the two Botanical Powder non-detections were non-detected under every condition, so concordance among qPCR conditions remained 100%. The comparison to compendial results for those samples is covered under Limit of detection.
Robustness
The internal amplification control amplified in every sample across all five matrices, both instruments, both extraction methods, and all inoculation levels. No systematic PCR inhibition was observed. Deliberate changes to method parameters were not part of this study; robustness here means tolerance to variation between matrices.
Matrix | Challenge it presents |
|---|---|
Energy shot | Aqueous liquid with organic acids, vitamins, caffeine |
Vitamin gummy | Semi-solid gelatin with sugars, colorants, vitamins |
Vitamin powder | Fine powder with mineral and vitamin components |
Botanical powder | Plant matrix with polyphenols, tannins, and other potential PCR inhibitors |
MCT oil | Hydrophobic lipid |
No matrix-specific modifications were needed beyond the pre-qPCR dilutions specified in the PathoSEEK® User Guide.
Equivalency
Non-inferiority to compendial plating was demonstrated at both inoculation levels tested using USP <1223> Approach 1 (paired samples, one-sided Z-test, δ = 0.20, α = 0.05).
Both organisms were co-inoculated into every test portion, so each portion gives one paired outcome; a portion counts as discordant if the two methods disagree on either organism. The six qPCR results per portion are repeated measurements of the same sample, so the analysis uses 50 independent test portions per level (10 per matrix across five matrices).
Level | Test portions | Positive rate, qPCR | Positive rate, compendial | Z-statistic | Critical value | Result |
|---|---|---|---|---|---|---|
Low | 50 | 0.960 | 1.000 | 5.77 | 1.645 | Pass |
High | 50 | 1.000 | 1.000 | ∞ (no discordant pairs) | 1.645 | Pass |
Source: Tables 8–10
USP <1223> recommends testing non-inferiority at a level where 50 to 75% of samples are positive, with at least 75 samples per procedure. This study used 50 test portions per level at levels with near-complete detection, so non-inferiority holds at the levels tested, and relative sensitivity near the detection threshold was not characterized.
Metric | Low level | High level | Overall |
|---|---|---|---|
Sensitivity, qPCR | 96% | 100% | 98% |
Sensitivity, compendial | 100% | 100% | 100% |
Relative trueness | 98% | 100% | 99% |
False positive rate | 0% | 0% | 0% |
Measurement-level figures, counting all six qPCR results per test portion. Source: Table 12.
Conclusion
The PathoSEEK® Staph and Pseudomonas Multiplex Detection Assay v2 with MaGiC Lysis met all five USP <1223> parameters and demonstrated equivalency to USP compendial plating for qualitative detection of S. aureus and P. aeruginosa in energy shot, vitamin gummy, vitamin powder, botanical powder, and MCT oil matrices.
Related Resources:
Frequently asked questions
Is PathoSEEK® equivalent to USP / plating for S. aureus and P. aeruginosa in dietary supplements?
In this study, the assay demonstrated non-inferiority to USP <2022> compendial plating (harmonized with USP <62>) at both low and high inoculation levels across five supplement matrices, evaluated against USP <1223>.
Which supplement matrices were tested?
Energy shot, vitamin gummy, vitamin powder, botanical powder, and MCT oil. They span aqueous liquids, gelatin semi-solids, powders, plant material, and oils.
Did any matrix show lower detection than compendial methods?
Botanical powder detected 8 of 10 low-level test portions versus 10 of 10 by plating. Both organisms were missed in the same two portions, which carried 1 to 4 CFU per test portion. At 10 samples, the report treats the difference as inconclusive (dPOD 95% CI −0.54 to 0.14) and notes that a sample-level matrix effect cannot be ruled out.
Does the assay work with automated extraction and different qPCR instruments?
Yes. Results were concordant across manual and automated (Myra) extraction and across the Bio-Rad CFX Opus and BioMolecular Systems Mic thermocyclers.
Does this study establish a formal limit of detection?
No. It confirmed detection at the levels tested, with low-level spikes ranging from 2 to 21 CFU per test portion depending on matrix and organism. A per-matrix LOD₅₀/LOD₉₅ determination was outside its scope.
If you have questions, please complete our “Contact Us” form and a Medicinal Genomics representative will contact you with details about how to get started with PathoSEEK® for Dietary Supplements.